HydraMist

Simultaneous Cold Vapor/Pneumatic Nebulization Spray Chamber

The Glass Expansion HydraMist is a sensitive, simple-to-use spray chamber for Inductively Coupled Plasma (ICP) that allows simultaneous operation of both conventional pneumatic nebulization and cold-vapor/hydride generation.

HydraMist Spray Chamber

Introduction

The Glass Expansion HydraMist is a sensitive, simple-to-use spray chamber for Inductively Coupled Plasma (ICP) that allows simultaneous operation of both conventional pneumatic nebulization and cold-vapor/hydride generation. Cold vapor generation can provide more than 10-fold improvement in sensitivity on ICP for cold vapor forming elements such As, Sb, Se, Tl and Hg. The generation of volatile species of these elements results in increased analyte loading of the analytical plasma giving lower detection limits.

The design of the HydraMist spray chamber is based upon Glass Expansion’s industry-standard cyclonic spray chamber, giving excellent sensitivity and short-term analytical precision with fast washout. The HydraMist spray chamber features a secondary inlet port that mixes the aerosolized sample and liquid reductant inside the spray chamber for rapid conversion of the As, Sb, Se, Tl, and Hg analytes into volatile hydride species. The unique drain design ensures fast, complete removal of waste from the spray chamber, eliminating excess hydrogen build-up that causes sample reflux degrading analytical precision.

Product Features

  • The same outstanding short-term analytical precision and washout as other Glass Expansion cyclonic spray chambers.
  • Fast and complete vapor phase formation of volatile As, Se, Sb, Tl and Hg species for the best detection limits in hydride generation mode.
  • A unique drain design to eliminate hydrogen build-up and sample reflux that degrades short-term precision.
  • Economic, just replace your current spray chamber and keep your existing nebulizer.
  • Improve productivity by analyzing non-hydride forming elements and cold vapor elements simultaneously, avoiding system shutdowns to change over between the hydride generator accessory and conventional pneumatic nebulization.
HydraMist Kit for Thermo
Above Image. HydraMist Kit for Thermo Scientific™ 7600 DUO ICP-OES

Product Benefits

With a simple setup and fast changeover, the HydraMist Spray chamber can be operated in one of 3 different modes. The simple mode provides (figure 1.) simultaneous hydride and pneumatic nebulization mode with 5-fold improvements for detection limits of the hydride-forming elements while maintaining analytical performance for non-hydride elements.

Sensitive simultaneous mode
Figure 1. Sensitive simultaneous cold vapor/pneumatic nebulization mode

With the addition of an extra sample input and t-piece (figure 2.), the HyrdraMist provides a sensitive, simultaneous cold vapor and pneumatic nebulization mode with more than 10-fold improvements in detection limits for the cold vapor elements without compromising the performance of non-hydride forming elements. Finally, it can be used as a conventional cyclonic spray chamber with pneumatic nebulization.

Simple simultaneous mode
Figure 2. Simple simultaneous cold vapor/pneumatic nebulization mode

Measured 3σ detection limits (in µg/L) with an Agilent Technologies 5100 SVDV in axial mode at 1.4 kW RF power and 20-second integration time.

Elements λ
Pneumatic Nebulization µg/L
Sensitive simultaneous hydride/pneumatic nebulization mode µg/L Simple simultaneous hydride/pneumatic nebulization mode µg/L
As 188 3.7 0.2 0.5
Hg 194 1.2 0.07 0.1
Sb 206 3.6 0.2 0.4
Se 196 2.9 0.2 0.5
Cd 214 0.1 0.1 0.1
Co 238 0.4 0.5 0.7
Cr 267 0.5 0.3 0.3
Cu 327 0.5 0.6 0.6
Fe 238 0.4 0.3 0.4
Mn 257 0.05 0.04 0.03
Mo 202 0.5 0.7 0.7
Ni 231 0.7 1.0 1.0
Pb 220 1.9 2.3 2.6
Tl 190 2.1 0.2 0.2
Zn 213 0.2 0.2 0.2

HydraMist Modes

HydraMist Modes

Ordering Information

Please select the kit to match your ICP from the following:

Part Number HydraMist Instrument Compatibility
KT-1157 Agilent® 5100, 5110, 5800, 5900
KT-1157 Analytik Jena® PQ 9000
KT-1157 PerkinElmer® Avio 200, 500
KT-1162 PerkinElmer® Optima 4000/5000/7000 DV
KT-2742 PerkinElmer® Avio with HTS System
KT-1162 PerkinElmer® Optima 4000, 5000, 7000 DV
KT-1162 PerkinElmer® Optima 8000, 8300 DV
KT-1157 Shimadzu® 7000, 8000, 9000, 9800 Series
KT-1160 Spectro™ Arcos II SOP/EOP and Blue EOP/SOP
KT-1160 Thermo Scientific™ PRO Duo & Radial
KT-1179 Thermo Scientific™ Duo 6000, 7200, 7400
KT-1156 Thermo Scientific™ Duo 7600

For all enquiries or more information:

Contact Glass Expansion